IP Water Ingress Test Equipment
A stainless-steel test solution compliant with
IP water ingress standards, with closed-loop water recirculation, adjustable
flow rate, and an integrated turntable.
It runs the full IPX1/X2/X3/X4/X5/X6 ingress test range. Test fittings can easily be swapped by hand to match the test being run.
The machine has two main parts: the test chamber and the standoff tube.
The test chamber is built from stainless chrome sheet. It consists of a watertight reservoir and a turntable for mounting the test specimen. Table diameter (Ø350–Ø500 mm) is set to customer specification. Interior lighting and a plexiglass cover allow the test to be observed from outside.
The standoff tube is 400 mm in diameter and 2,5 m long. A flow meter and a manual valve set the water flow. A crank handle adjusts the height to suit the product under test. It is used for IPX5/X6 water jet tests.
Device moves and locks into place with little effort.
Technical Specifications
Chamber Dimensions: 1000x1000x1500 mm (W × L × H)
Standoff Tube Dimensions: 580x2500x1340 mm (W × L × H)
Gearmotor Power: 400 V / 0,18 kW / 24 rpm
Pump Power: 220 V / 1,1 kW / 0–120 L/min
Electrical Connection: 380 V / 1,3 kW (220 V optional)
IP Test Equipment Technical Drawings


** All dimensions are given in mm.
** Chamber dimensions can be customized to meet customer requirements.
General Description
IPX1 TEST CONDITIONS
WATER USED: Fresh water or pure water must be used during the test.
DURATION: Test duration must be 10 minutes.
WATER FLOW RATE: 0,25 L/min (To reach this flow rate, the flow meter is monitored and adjusted via the valve.)
ROTATION SPEED: 1 rpm
PRESSURE and TEMPERATURE: Atmospheric pressure / Room temperature
DISTANCE: The distance between the drip box and the test specimen must be at least 200 mm.
IPX2 TEST CONDITIONS
WATER USED: Fresh water or pure water must be used during the test.
DURATION: Test duration must be 10 minutes. (4×2,5 min)
WATER FLOW RATE: 0,75 L/min (To reach this flow rate, the flow meter is monitored and adjusted via the valve.)
ROTATION SPEED: Fixed at 15° tilt, no rotation
PRESSURE and TEMPERATURE: Atmospheric pressure / Room temperature
DISTANCE: The distance between the drip box and the test specimen must be at least 200 mm.
IPX3 TEST CONDITIONS
WATER USED: Fresh water or pure water must be used during the test.
DURATION: Test duration must be 5 minutes.
WATER FLOW RATE: 10 L/min ±5% (To reach this flow rate, the instantaneous flow is monitored on the flow meter and adjusted via the valve.)
ROTATION SPEED: 10 rpm (Optional)
TUBE ROTATION ANGLE: ±60° (120° total)
PRESSURE and TEMPERATURE: Atmospheric pressure / Room temperature
DISTANCE: The distance between the oscillating tube and the test specimen must be approximately 300 mm.
IPX4 TEST CONDITIONS
WATER USED: Fresh water or pure water must be used during the test.
DURATION: Test duration must be 10 minutes.
WATER FLOW RATE: 10 L/min ±5% (To reach this flow rate, the instantaneous flow is monitored on the flow meter and adjusted via the valve.)
ROTATION SPEED: 10 rpm (Optional)
TUBE ROTATION ANGLE: ±120° (240° total)
PRESSURE and TEMPERATURE: Atmospheric pressure / Room temperature
DISTANCE: The distance between the oscillating tube and the test specimen must be approximately 300 mm.
IPX5 TEST CONDITIONS
WATER USED: Fresh water or pure water must be used during the test.
DURATION: Test duration must be 3 minutes.
WATER FLOW RATE: 12,5 L/min ±5% (To reach this flow rate, the instantaneous flow is monitored on the flow meter and adjusted via the valve.)
ROTATION SPEED: 20 rpm
PRESSURE and TEMPERATURE: Atmospheric pressure / Room temperature
DISTANCE: The distance between the nozzle and the test specimen must be between 2.5 m and 3 m.
IPX6 TEST CONDITIONS
WATER USED: Fresh water or pure water must be used during the test.
DURATION: Test duration must be 3 minutes.
WATER FLOW RATE: 100 L/min ±5% (To reach this flow rate, the instantaneous flow is monitored on the flow meter and adjusted via the valve.)
ROTATION SPEED: 20 rpm
PRESSURE and TEMPERATURE: Atmospheric pressure / Room temperature
DISTANCE: The distance between the nozzle and the test specimen must be between 2,5 m and 3 m.
Test Evaluation Criteria
· After the test, the exterior of the equipment is should be dried and the interior should be inspected.
· No water or moisture of any kind should be present inside the device.
· The electronic components of the device must remain fully functional.
· The device must not show any physical deformation due to pressure.